Development of a Novel Multiplex Bead-based Assay for Measuring Autoantibodies on Flow Cytometric Platform.

Chauhan R, Gupta N, Tiwari AK, Raina V, Nandi SP

Abstract

Background: Autoantibodies (AAbs) are important biomarkers for the diagnosis of Autoimmune Diseases (ADs). The detection of AAbs performed by current methods (indirect immunofluorescence test (IIFT)/Immunoblot (dot/line)/enzyme-linked immunosorbent assay ELISA) which have limitations in terms of performing multiple assays to arrive at laboratory diagnosis. We validated a novel multiplex bead-based assay (NMBA) that could quantify five common antibodies, simultaneously, on a flow-cytometry platform.Methods: A total of five recombinant antigens (SS-A Ro60, CENP B, RNP 70, Scl 70 and Histones) were covalently coupled onto beads and tested using known positive sera (positive for AAbs) and analyzed using flow cytometer.Results: The sensitivity, specificity, Positive Predictive Value (PPV) and Negative Predictive Value (NPV) were obtained for each antigen, analyzed by both assays (NMBA and IIFT). It showed comparable or higher values for the NMBA. The Spearman's rank correlation coefficient (Rho) were ≥ 0.97, (P < .05), indicating that multiplexing of the five autoantigens did not alter the results obtained when antigens were tested individually. The mean intra-assay precision measured by coefficient of variation (CV) was7.56 ± 1.6% and the mean inter-assay CV was 10.03 ± 1.34%. The time taken from sample receipt to reporting of results was 90 minutes in NMBA as compared to 150 minutes of IIFTConclusion: The NMBA could quantitatively measure antibodies against five autoantigens, simultaneously in patient's sera. The assay is faster, objective, reproducible, requires low sample volume, and stable. Moreover, the flow cytometer in diagnostic laboratory settings for hematological and transplant immunology tests, can also be used for testing AAbs.

Keywords
Autoantibodies autoimmune diseases bead-based flow-cytometric multiplex
Article Info
Journal
Immunological investigations
Abbr.
Immunol Invest
ISSN
1532-4311
Published
2020-12-07
Language
English
Country/Region
England
NLM ID
8504629
Analysis Services
Analysis Services

Contact

No. 2 Wenbo Road, Zhangqiu District, Jinan, Shandong

Qilu Normal University · Genelibs Bioinformatics Lab

750 Shunhua Rd, Jinan

2F, Bldg F, University Science Park

Tel: 0531-88819269

WeChat Official Account

Follow our WeChat subscription account for real-time updates and the latest in medical and biological research.


Business Email

E-mail: product@genelibs.com